HIOKI 3290-10 Clamp On AC/DC HiTester All the Functions You Need for Measurement at DC or 1Hz and Up
Choice of three sensors (Example combinations)
3290/3290-10 +9691 : Measure up to 100A
3290/3290-10 +9692 : Measure up to 200A
3290/3290-10 +9693 : Measure up to 2000A
Choice of measurement methods
DC (for battery measurement)
AC+DC RMS (for full-/half-wave rectification measurement)
AC RMS (for current distortion measurement)
PEAK (for peak value measurement) of inrush current, etc.)
Choice of output (Simultaneous output)
Effective value output, frequency output, waveform output
3290-10 Functions
Current integral measurement
(obtain polarity-specific integrated DC values)
Operating time/duty measurement
Measurement method : DC, AC+DC (True RMS), AC (True RMS)
Peak value display : Polarities independently (±peak value of waveform, in DC mode)
Separate AC from DC output : None
Output (terminal 1) : Current waveform or true RMS value (2 V / setting range)
Output (terminal 2) : Current integral value (1 VDC / setting range)
Integral current measurement : ±, +, - integral, average value
Rate measurement : Operating time rate, total measurement time
Timer settings : 99 hr, 59 min. (for repeated up to 20 times)
Data storage : Peak, integral value, operating time rate, or other
Basic accuracy at DC current : 20.00/ 100.0A (using the 9691 sensor): ±1.3 % rdg. ±0.10 A/ ±0.5 A
20.00/ 200.0A (using the 9692 sensor): ±1.3 % rdg. ±0.10 A/ ±0.5 A
200.0/ 2000A (using the 9693 sensor): ±1.8 % rdg. ±1.0 A/ ±5 A
Basic accuracy at AC current
(45 to 66 Hz) 20.00/ 100.0A (using the 9691 sensor): ±1.3 % rdg. ±0.08 A/ ±0.3 A
20.00/ 200.0A (using the 9692 sensor): ±1.3 % rdg. ±0.08 A/ ±0.3 A
200.0/ 2000A (using the 9693 sensor): ±1.3 % rdg. ±0.8 A/ ±3 A
Voltage measurement : None
Frequency measurement : 1.00Hz to 1000Hz, Basic accuracy: ±0.3% rdg. ±1dgt.
Crest factor : 2.5 or later, or rated peak value of sensor
Frequency characteristic : DC, or 1 Hz to 1 kHz (accuracy guaranteed range)
At monitor output: based on the using sensor
Other functions : Selectable of responce time, filter (fc=550 Hz)
Power supply : LR6 (AA) alkaline batteries × 4, Continuous use : 22 hr, or AC adapter 9445-02/-03 (The 3290-10 only: external power supply possible)
Dimensions, mass 155 mm (6.10 in) W × 98 mm (3.86 in) H × 47 mm (1.85 in) D, 545 g (19.2 oz)
Supplied accessories : Carrying strap ×1, LR6 (AA) alkaline batteries × 4